Principles and Applications of Fourier Transform Infrared Spectroscopy (FT-IR)


I. Instrument Structure and Principle

1. Definition

Mid-infrared light refers to radiation with wavelengths ranging from 2.5 to 25 μm (or 4800 to 400 cm⁻¹). When this light is directed onto a sample, it can be absorbed, transmitted, reflected, scattered, or used to excite fluorescence (i.e., the Raman effect). After absorbing mid-infrared light, molecules undergo changes in their vibrational and rotational states, resulting in an infrared absorption spectrum.

An instrument that generates mid-infrared light and records infrared absorption spectra is called a mid-infrared spectrometer.

When a sample is irradiated with infrared light whose frequency varies continuously, molecules absorb radiation at certain frequencies, causing a net change in their dipole moments due to vibrational or rotational motions. This results in transitions of molecular vibrational and rotational energy levels from the ground state to excited states, thereby reducing the intensity of transmitted light corresponding to these absorption regions. By plotting the percentage transmittance of infrared light against wavenumber or wavelength, one obtains the infrared spectrum.

2. The necessary conditions for generating infrared spectra

(1) The frequency of infrared radiation must be comparable to the frequency of molecular vibrations in order to provide the energy required for molecular vibrational energy level transitions, thereby producing an absorption spectrum.

(2) During vibration, only molecules that can induce a change in molecular dipole moment are capable of producing an infrared absorption spectrum.

3. Various vibration modes and energy

Molecular vibrational modes are classified as:

Stretching vibration – Symmetric stretching vibration

---- Asymmetric Stretching Vibration

Bending vibration – In-plane bending vibration – Shear vibration

----Planar Swing

----Out-of-plane bending vibration----Non-planar oscillation

----Torsional Vibration

IR spectrum of carbon dioxide:

4. Representation of Infrared Spectra

Generally refers to the mid-infrared (vibrational energy level transitions).

Mid-infrared absorption spectrum:

Horizontal coordinate: Wavenumber is the reciprocal of wavelength.

400–4000 cm⁻¹; indicates the position of the absorption peak.

Y-axis: Transmittance (T%), indicating the absorption intensity. A lower T value indicates better absorption; therefore, the trough of the curve represents a favorable absorption band.

Information obtainable from the spectrum:

(1) The position of the absorption peak (absorption frequency)

(2) The intensity of absorption peaks is commonly indicated by the following terms: vs (very strong), s (strong), m (medium), w (weak), vw (very weak), b (broad), sh (sharp), v (variable).

(3) The shape of the absorption peak (sharp peak, broad peak, shoulder peak)

5. Construction and Principle of FT-IR Spectrometer

Infrared light source (Globr) → Interferometer → Sample chamber → Detector (HgCdTe) (combined light source)

Interferometer: The heart of the infrared spectrometer, consisting of a fixed mirror, a movable mirror, and a beam splitter (KBr).

At the output of the interferometer, two beams of light with a path difference interfere with each other and then proceed to the sample.

Fixed mirror: reflects the light beam.

Moving mirror: The beam is modulated by controlling the scanning speed of the scanner with a laser, generating interference light.

Beam splitter: transmitted beam.

Detector: The interference signal is amplified, filtered, and digitized, then subjected to a mathematical transformation (Fourier transform) to produce a spectrum.

II. Common Preparation Methods for Infrared Samples

1. Tablet Pressing Method

Grind KBr (100–200 mg) together with the solid sample (1–2 mg) in an agate mortar to obtain a fine powder with particle sizes in the micrometer range. Then, using a specialized pelletizing device, compress the mixture into a transparent pellet approximately 13 mm in diameter and about 1 mm thick for subsequent analysis.

In addition to KBr, the diluents used in the pellet method also include NaCl, CsI, and polyethylene powder.

2. Paste Method

The finely ground solid sample powder (10 mg) is mixed with a small amount of fluorinated kerosene (which exhibits no infrared absorption in the 4000–1300/cm region) or liquid paraffin (which exhibits no infrared absorption in the 1300–400/cm region) to form a paste, which is then applied onto a salt plate or an insoluble window for analysis.

The paste method can eliminate interference from water peaks (3400/cm, 1630/cm); alternatively, adding a few drops of heavy water to the sample can also remove the interference of water peaks on the sample signal.

When studying samples containing -OH and -NH2 groups, to eliminate interference from moisture in KBr, you can use polytetrafluoroethylene powder as a tabletting carrier or adopt the paraffin paste method.

3. Coating Method

To perform the analysis, simply drop or spread a liquid sample onto a salt plate or window plate to form a liquid film. For certain solid polymers, thin films suitable for analysis can also be obtained by melting and coating, hot-pressing into films, or casting from solutions.

Common salt crystals: KBr, NaCl.

Common water-insoluble window materials: CaF2, BaF2, and KRS-5 (thallium bromoiodide).

4. Liquid Pool Method

Liquid samples are analyzed using liquid cells of various specifications and materials. The cell thickness ranges from 0.01 to 1 mm, and the cell materials include KBr, NaCl, and others. When the sample is an aqueous solution, KRS-5 or AgCl windows—materials that are insoluble in water—can be selected. Since liquid cells are not very convenient to assemble and disassemble, they are used only when infrared spectroscopy is employed as a quantitative analytical method.

5. Gas Sample Analysis

FT-IR technology can be used for the direct analysis of gas samples.

Various gas cells (including conventional gas cells, small-volume gas cells, long-path-length gas cells, pressurized gas cells, high-temperature gas cells, and low-temperature gas cells) and vacuum systems are essential auxiliary equipment and accessories for gas analysis.

III. Applications of Infrared

1. Applications of infrared

(1) Distinguish conformational isomers by leveraging the differences in solid- and liquid-phase spectra.

(2) Identify the functional groups present in the compound based on the characteristic absorption peaks.

(3) Determine the purity of the sample. If the sample contains more than 5% impurities,

(4) The sharpness of the spectral absorption peaks decreases, and the number of absorption peaks increases.

(5) Quantitative Analysis: Using an internal standard peak or selecting two absorption peaks, construct a calibration curve by plotting the ratio of peak intensities against sample concentration.

(6) By observing changes in the intensity of a specific spectral peak, one can track the progress of a chemical reaction.

(7) Identify compounds (polymorphs, homologs, optical isomers, and geometric isomers) by leveraging differences in solid-phase spectra.

2. Division of Infrared Spectroscopy:

(1) 4000–2500 cm⁻¹: This is the stretching vibration region for X-H single bonds. X can be an atom of O, N, C, or S.

(2) 2500–2000 cm⁻¹: This region corresponds to the stretching vibrations of triple bonds and conjugated double bonds.

(3) 2000–1500 cm⁻¹: This region corresponds to the stretching vibration of double bonds.

(4) 1500–600 cm⁻¹: This region primarily provides information on C-H bending vibrations.

The unsaturated C-H stretching vibration, appearing above 3000/cm, can be used to determine whether a compound contains unsaturated C-H bonds, such as -C=C-H. Meanwhile, the absorption peak corresponding to the double-bond stretching vibration appears between 2000 and 1500/cm.

The C=O bond in an acid anhydride gives rise to two peaks due to vibrational coupling. The relative intensities of these two peaks can be used to determine whether the acid anhydride is linear or cyclic: in a linear acid anhydride, the intensities of the two peaks are approximately equal, with the higher-wavenumber peak being only slightly stronger than the lower-wavenumber peak; whereas in a cyclic acid anhydride, the lower-wavenumber peak is actually stronger than the higher-wavenumber peak.

When there are four or more CH2 groups on a single chain, the 720 ± 10 cm⁻¹ band corresponds to the wagging vibration of the CH2 groups.

Share to:

Related News


Application of High-Performance UV Resins in Vacuum Electroplating Coatings

Plastic materials are widely used, easy to mold, and lightweight. Mold injection molding offers excellent dimensional stability and can produce intricate surface textures that are difficult to achieve with any metal material. After electroplating, the surface becomes metallic, giving it a high-quality metallic finish. These plastics can serve decorative purposes as well as provide protection against electromagnetic interference, reflection, or anti-reflection effects.


Principles and Applications of Gas Chromatography

Gas chromatography is a major scientific and technological breakthrough that emerged in the 1950s. It is a new separation and analytical technique that has found wide applications in industry, agriculture, national defense, construction, and scientific research.


Design Concept and Application & Construction Control of UV Coating Formulations for Enclosures

In different applications, UV coatings play varying roles, and there are significant differences in their physicochemical properties and the choice of coating types. Generally, the appropriate UV coating is selected based on the application method and conditions, and professional technical personnel should provide guidance.